Hypatia Models 308 and 309 Performance Verification Procedure
Warning:
Potentially lethal line voltages of as much as 132 Volts rms are
present on exposed terminals within this instrument. Opening the
instrument cabinet is not necessary and is not recommended for
the performance verification procedure. If repair service is required
but you are unfamiliar with the precautions and techniques necessary
for servicing instruments with exposed line voltages, do not attempt
to service the instrument yourself. Instead, refer service to
appropriately experienced technicians.
Warning:
This instrument produces exposed high currents as a normal part
of its function. High current shorts through rings, wrist bands
or other conductive apparel can cause serious thermal burns which
may result in permanent injury. Operators are advised to remove
all conductive apparel, especially from the hands and wrists,
prior to any operation or service of this instrument.
No Adjustments: There are no hardware adjustments within
this instrument. There is no need to open the cabinet to perform
a performance verification. If the instrument fails any performance
verification step, replacement of the associated components or
cleaning of the instrument will be required. Contact Hypatia for
service documentation.
Required Probes and Network Cable: Model 109 Kelvin
sense probes are required. (If unavailable, they can be easily
built: Terminate four conductor telephone type flat cable with
a telephone type modular connector on one end and solder all four
conductors to a small alligator clip on the other end. A 1.5 meter
length is suggested for each probe, but not required.)
Two alligator clip probes (Model 101 or 105) are recommended,
but are not required so long as a stout means of attaching two
Model 100 series power probes to a power resistor bank and the
100 milliohm cabinet mounted check resistor can be achieved.
One serial network cable with a Mini DIN-8 connector on one
end and a suitable serial connector for a Macintosh or a PC on
the other end (provided to all customers).
Required Software, Computer, Test Instruments and Fixtures:
Current
Model 309 Utility software and Model
309 Operating Software (also used in the model 308)
A 32 bit capable "Wintel" compatible desktop computer with an EIA-232-E (aka RS-232) serial port.
One RMS voltmeter with at least .2% accuracy and at least 1
mV resolution.
One DC voltmeter with at least .04% accuracy and at least 1 mV
resolution at 2.50 and 4.096 Vdc.
One digital counter with at least .02% accuracy and a range of
at least 16.8 MHz.
One precision power resistor bank, with sense connection points,
totaling at least 100 milliohm with at least .2% accuracy, and
with taps if verification at readings of less than 100 milliohm
are desired. This resistor bank must have a current handling capacity
at least meeting the capacity and duty cycle limits of the Models
308 and 309.
1. Using the Model
309 Utility software, install the latest version of the Model 309 Operating Software
into the instrument (the Model 309 Operating Software is also
used in the Model 308). This software is available free on the
web at www.Hypatia.com.
If necessary see the owner's manual for installation instructions,
also on the web. Note: Installation of the latest Operating Software
is mandatory - no exceptions. These are safety mission
instruments. Failure to install the latest Operating Software
is negligent and may create a serious legal liability. Hypatia
specifically prohibits certification of instruments using any
obsolete version of the Operating Software, and does not warrant
the accuracy or performance of any instrument so serviced. Hypatia promptly reports third party service
quality failures to instrument owners whenever we become aware
of them.
2. Set the Model 309 "main power" switch to the off
position. Rotate the "current adjust" knob to its full
counter clockwise position.
3. Connect an IEC power cord to the rear panel IEC receptacle
of the Model 309 and to a power source meeting the specifications
printed above the IEC receptacle.
4. Set the Model 309 "main power" switch to the on position
and allow five minutes to establish near thermal equilibrium.
5. Secure two Model 100 series power probes, preferably alligator
clip types (Model 101 or 105), to the front panel connectors of
the instrument as described in the operation section of the manual,
but leave the modular connectors unconnected. Connect the Kelvin
sense probes (Model 109) to the modular receptacles. Connect the
power probe ends to a 100 milliohm precision power load. Connect
the Kelvin sense probe ends to the sense taps of the precision
load, being certain that the right Kelvin sense line is connected
to the same side of the load as the right power probe, and similarly
for the left sense line and probe. Verify that the Kelvin connectors
do not directly physically touch the power probe connectors.
6. Set the main readout to the milliohm mode. If enabled, disable
the test timer and all alarms.
7. Push the "start test" switch, then slowly increase
the test current to 2.5 Amps. Record the Model 309 impedance reading.
Repeat for 10.0, 17.5, 25, 30 and 60 Amps. Perform the 60 Amp
test quickly so as not to violate duty cycle limits. Push the
"start test" switch again to stop the test. Rotate the
"current adjust" knob to its full counter clockwise
position. Compare the results to specifications.
8. If impedance measurement performance checks for load values
other than 100 milliohms are desired, repeat step six for each
value.
9. If altered, reset the load bank to 100 milliohms. Connect an
RMS voltmeter to the sense taps of the precision load. Set the
Model 309 main readout to the current mode.
10. Push the "start test" switch, then slowly increase
the test current until the RMS voltmeter reads 250 mV, which equates
to 2.5 Amps. Record the Model 309 digital current reading. Vary
the current adjust knob slightly to confirm that bar graph threshold
crossing matches the digital reading. Repeat for 10.0, 17.5, 25,
30 and 60 Amps. Perform the 60 Amp test quickly so as not to violate
duty cycle limits. Push the "start test" switch again
to stop the test. Rotate the "current adjust" knob to
its full counter clockwise position. Compare the results to specifications.
11. Set the Model 309 main readout back to the milliohm mode.
Establish a connection between the probe's measurement ends and
the 100 milliohm check resistor on the back of the instrument.
Use the Kelvin sense lines, being certain that the right Kelvin
sense line is connected to the same side of the 100 milliohm resistor
as the right power probe, and similarly for the left sense line
and probe. Verify that the Kelvin connectors do not directly physically
touch the power probe connectors.
12. Push the "start test" switch and rotate the "current
adjust" knob clockwise until a current of 2.5 Amps is reached.
Record the impedance reading. Push the "start test"
switch again to stop the test. Rotate the "current adjust"
knob to its full counter clockwise position. Compare the reading
to the Model 309's reading of the 100 milliohm precision load
taken earlier at 2.5 Amps to determine the value of the cabinet
mounted check resistor by error subtraction. The result should
be 100 milliohms ± 1%.
13. [This step not applicable to the Model 308.] Check operation
of the milliohm, high current and low current alarms. The set
points of the alarms may be reviewed (and changed if necessary)
by cycling through the adjust menu ("adjust select"
button). Execute tests which exceed each alarm's threshold and
verify that an audio alert sounds and the associated alarm light
flashes. Stop the last test and rotate the "current adjust"
knob to its full counter clockwise position.
14. Remove all of the probes except the left Kelvin sense probe
from the Model 309. Turn the Model 309 on its side and note the
two probe ports on the bottom near the front of the instrument.
15. Connect the reference terminal of a DC voltmeter to the alligator
clip end of the left Kelvin sense probe. Insert the other voltmeter
probe into the probe port marked "2.500 V" or "4.096
V". Compare the reading to the ideal of 2.500 V or 4.096
V (2E12 mV) and verify that accuracy is within .2%.
16. Connect the reference terminal of a digital counter to the
alligator clip end of the left Kelvin sense probe. Insert the
other counter probe into the probe port marked "16.777 MHz".
Compare the reading to the ideal of 16.777216 MHz (2E24 Hz) and
verify that accuracy is within .1%.
17. Set the "main power" switch to the off position,
turn the Model 309 to its normal, upright position, and remove
the left Kelvin sense probe and the power cord from the instrument.
18. Check that the two lower rear #15 torx screws which secure
the top cabinet half are tight and that there is a lockwasher
located between the cabinet halves on these screws to insure proper
grounding of the upper cabinet half. The edges of the lockwashers
are visible from the bottom of the instrument.
This Completes the Performance Verification Procedure.